EasyManua.ls Logo

Steris Vision 1321 - Page 129

Steris Vision 1321
210 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 Routine Maintenance
118
920515770_J
Vision® 1300 Series
Operator Manual
Table 7-1. Routine Maintenance Checklist (continued)
Usage and utility conditions may require more or less frequent inspections.
Subsystem
4.0 4X PER YEAR
4.1
Inspect cycle printouts and PV reports for signs of trouble.
Preparation
4.2
Verify sump under floor sections for debris. Use a hose to clean if necessary.
Chamber
4.3
Verify around interior of reservoir(s) and also reservoir(s) cover to see if any hard water
deposits are visible. Brush off visible hard water deposits before DECONTAM cycle (see
Section 7.6, Decontam Cycle).
IMPORTANT: Procedure to be performed by technician or biomed personnel only.
Mechanical Core
4.4
Verify pickup tube is not clogged, cracked or distorted. If damaged, contact STERIS* or
Customer ’s Maintenance Service.
Chemicals
4.5
Verify operation of chemical empty switches. If damaged, contact STERIS* or Customer’ s
Maintenance Service.
Chemicals
4.6
Verify washer/disinfector has proper labels (caution, warning, etc.).
Final Test
5.0 2X PER YEAR
5.1
Test emergency stop cables for proper operation (see Section 7.11, Emergency Stop
Cables).
Safety Feature
5.2
Test Emergency Stop pushbuttons for proper operation (see Section 7.12, Emergency Stop
Pushbuttons).
Safety Feature
5.3
Test Lock/Unlock key selector switch for proper operation. Safety Feature
6.0
AS REQUIRED
6.1
Inspect for proper operation of printer. Verify printout for darkness, missing dots, etc.
Replace ribbon cartridge if necessary (see Section 7.8, Changing Printer Paper Roll and Ink
Cartridge (Accessory)).
Printer
6.2
Change printer paper roll (see Section 7.8, Changing Printer Paper Roll and Ink Cartridge
(Accessory)).
Printer
6.3
Change chemical container (see Section 7.7, Chemical Container Replacement).
Chemicals
* Contact STERIS for this Service.

Table of Contents