8 Troubleshooting
182
920515770_J
Vision® 1300 Series
Operator Manual
8.53 Warning# 5 Cold Water Pressure Below Requirement
Condition
Wash reservoir is too long to fill.
General Troubleshooting
Additional Information
Cold water valve is not fully open.
1. Verify if washer/disinfector cold water valve is fully
open.
2. If not, fully open cold water valve.
3. If situation reoccurs, contact STERIS.
Washer/disinfector cold water strainer is clogged.
1. Clean washer/disinfector cold water strainer.
2. If situation reoccurs, contact STERIS.
Cold water supply minimum pressure does not meet
requirement.
1. Contact STERIS.
8.54 Warning# 6 Cold Water Pressure Above Requirement
Condition
Wash reservoir fills too quickly.
General Troubleshooting Additional Information
Cold water pressure is too high.
1. Contact STERIS.
8.55 Warning# 7 Cycle Parameters Have Been Reset to Factory Default Values
Condition
At power up, cycles data file was detected as corrupted.
General Troubleshooting Additional Information
Cycles data file is corrupted.
1. Supervisor needs to reprogram reservoirs and
cycles parameters in Supervisor mode or through
Modify cycle screen.