7 Routine Maintenance
154
920515770_J
Vision® 1300 Series
Operator Manual
.
4. An alarm should be displayed on the control and air is released from door mechanism. Display shows:
5. Use key to turn LOCK/UNLOCK selector to UNLOCK position to unlock washer/disinfector.
6. Acknowledge alarm and resume washer operation.
7. Repeat procedure to test emergency stop cable on other side.
7.12 Emergency Stop Pushbuttons
Verify Emergency Stop pushbuttons as follows: