EasyManua.ls Logo

Steris Vision 1321 - Load comes out dirty at cycle end

Steris Vision 1321
210 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8 Troubleshooting
190
920515770_J
Vision® 1300 Series
Operator Manual
8.77 Load comes out dirty at cycle end
Condition
Load comes out dirty at cycle end.
General Troubleshooting Additional Information
Wrong type of detergent is used.
1. Follow detergent manufacturer’s recommendations
for quantity of detergent to be used.
2. If situation reoccurs, contact STERIS.
Traveler spray arms clogged.
1. Clean traveler spray arms (refer to Section 7.9,
Traveler Spray Arm Assemblies).
2. If situation reoccurs, contact STERIS.
Rotary spray arm assemblies on Vision CW Instrument
Washing Manifolded Rack clogged.
1. Clean rotary spray arms (refer to Section 7.10, Vision
CW Instrument Washing Manifolded Rack Spray Arm
Assemblies).
2. If situation reoccurs, contact STERIS.
Chamber debris screen clogged.
1. Clean debris screen (refer to Section 7.4, Debris
Screen).
2. If situation reoccurs, contact STERIS.
Wash phase too short.
1. Increase wash phase time in Supervisor mode
Modify cycle screen.
2. If situation reoccurs, contact STERIS.
Wash or thermal rinse reservoir recycling count too high.
1. Lower wash or thermal rinse reservoir recycling
count in Supervisor mode.
2. If situation reoccurs, contact STERIS.

Table of Contents

Related product manuals