8 Troubleshooting
191
Operator Manual
Vision® 1300 Series
920515770_J
8.78 Manifolded rack spray arms not turning
Condition
Manifolded rack spray arms not turning.
NOTE: Problem applies if Instrument Washing option is present.
General Troubleshooting
Additional Information
Clogged manifolded rack spray arms.
1. Clean spray arms (refer to Section 7.10, Vision CW
Instrument Washing Manifolded Rack Spray Arm
Assemblies).
2. If situation reoccurs, contact STERIS.
Excessive scale build up in manifolded rack spray arm
pivot.
1. Run a DECONTAM cycle with manifolded
instrument rack (refer to Section 7.6, Decontam
Cycle).
2. If situation reoccurs, contact STERIS.
8.79 No audible signal when alarm occurs, when cycle is completed, etc.
Condition
No audible signal when alarm occurs, when cycle is completed, etc.
General Troubleshooting Additional Information
Defective amplifier/speaker system.
1. Contact STERIS.
8.80 No display on touch screen(s)
Condition
No display on touch screen(s).
General Troubleshooting
Additional Information
Washer/disinfector needs to be restarted.
1. Shut washer/disinfector power OFF.
2. Restart washer/disinfector.
3. If situation reoccurs, contact STERIS.
Faulty touchscreen.
1. Contact STERIS.