5 Operating Instructions
79
Operator Manual
Vision® 1300 Series
920515770_J
... then, display shows Ready screen:
5. Resume washer/disinfector normal operation.
NOTE: Items must be reprocessed through a complete cycle (see Section 5.14, Typical Cycle Operation General,
for details on how to process a cycle).
5.30 Visual Inspection
IMPORTANT: Good Hospital Practice dictates all instruments be inspected for visible debris after processing in the
washer/disinfector. Any instrument with visible debris must be rewashed until clean and free of visible debris prior to terminal
processing. Failure to reprocess until all visible debris have been removed may impede the terminal processing.
1. Once items are cool, verify for any residue or soil.
2. Any item that has not been cleaned properly must be reprocessed through a complete cycle.