8 — Troubleshooting
162
920515770_J
Vision® 1300 Series
Operator Manual
8.11 Alarm# 9 Sump Draining/Recycling Malfunction
Condition
Problem in sump draining or water recycling system.
Alarm is triggered by...
• WATER RECIRCULATION LOW SPEED PRESSURE DETECTED input is activated for 120 consecutive seconds during
RETURN SUMP sequence.
General Troubleshooting
Additional Information
Problem in sump draining or water recycling system.
1. Press ALARM REPLY on touch screen to stop alarm
buzzer and acknowledge displayed alarm message.
2. Abort cycle.
3. If condition reoccurs, contact STERIS.
8.12 Alarm# 10 Water Recirculation Pressure Transmitter Defective
Condition
Water recirculation pressure transmitter wires are shorted or disconnected.
Alarm triggered by...
• WATER RECIRCULATION PRESSURE transmitter [PE1] wires are shorted or disconnected.
General Troubleshooting
Additional Information
Problem with water recirculation pressure transmitter
wires.
1. Press ALARM REPLY on touch screen to stop alarm
buzzer and acknowledge displayed alarm message.
2. Abort cycle.
3. If condition reoccurs, contact STERIS.