7 Routine Maintenance
137
Operator Manual
Vision® 1300 Series
920515770_J
7.9 Traveler Spray Arm Assemblies
Traveler Spray Arm Assemblies General
WARNING
PERSONAL INJURY HAZARD
Always verify door bridge is down before entering washer/disinfector chamber.
Keep hands and fingers away from door or any moving parts to prevent crushing.
WARNING
PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD
Before servicing or cleaning washer/disinfector, always turn LOCK/UNLOCK key selector
to LOCK position to prevent cycle operation. Keep the key on yourself.
WARNING
BURN HAZARD
Allow washer/ disinfector to cool down before performing any service on pump,
mechanical components or piping. Surfaces become very hot during washer/ disinfector
operation.
Inspect and Clean Traveler Spray Arm Assemblies
After last cycle of the day, clean traveler spray arm assemblies as follows:
1. Allow washer/disinfector to cool down before entering washer/disinfector chamber.
2. Press
on load side touch screen to open load side door.