7 — Routine Maintenance
138
920515770_J
Vision® 1300 Series
Operator Manual
3. Use key to turn LOCK/UNLOCK selector to position to lock washer/disinfector.
IMPORTANT: Retain key while inside chamber.
4. Enter chamber.
5. Use a fine wire (approximately wire gauge of a paper clip) to clean visible debris from spray arm jets by pulling
debris out of holes. If desired, spray jet cleaning needle (117072281) can also be used.
6. Exit chamber.
7. Use key to turn LOCK/UNLOCK selector to
position to unlock washer/disinfector.
8. Press