8 Troubleshooting
187
Operator Manual
Vision® 1300 Series
920515770_J
8.69 Excessive vapor exhausting from doors
Condition
Excessive vapor exhausting from doors.
General Troubleshooting
Additional Information
Door gasket is dirty.
1. Clean door gasket (refer to Section 7.5, Clean
Washer/Disinfector Exterior).
2. If condition reoccurs, contact STERIS.
8.70 Foam in chamber or inside reservoirs
Condition
Foam in chamber or inside reservoirs.
General Troubleshooting
Additional Information
Wrong type of detergent is used.
1. Use correct type of detergent (refer to washer
detergent specifications).
2. Follow detergent manufacturer’s recommendations
for quantity of detergent to be used.
Too much chemical injected during phase.
1. Lower injection rate.
Chemical pump needs calibration.
1. Contact STERIS.
8.71 Floor tilting system lifting too slowly, not lifting or stays in upper position
Condition
Floor tilting system lifting too slowly, not lifting or stays in upper position.
General Troubleshooting Additional Information
Load exceeds 600 lb (272 kg).
1. Verify load weight.
Air flow control improperly adjusted.
1. Contact STERIS.