8 Troubleshooting
181
Operator Manual
Vision® 1300 Series
920515770_J
8.50 Warning# 2 Pure Water Pressure Below Requirement
Condition
If Pure Water Line option is present: Pure water pressure measured by flowmeter is below requirements for 5
consecutive seconds.
General Troubleshooting
Additional Information
Pure water valve is not fully open.
1. Verify if washer/disinfector pure water valve is
fully opened.
2. If not, open washer/disinfector pure water valve.
3. If situation reoccurs, contact STERIS.
Washer/disinfector pure water strainer is clogged.
1. Clean washer/disinfector pure water strainer.
2. If situation reoccurs, contact STERIS.
8.51 Warning# 3 Hot Water Pressure Above Requirement
Condition
Hot water pressure measured by flowmeter is above requirements for 5 consecutive seconds.
General Troubleshooting Additional Information
Hot water pressure is too high.
1. Contact STERIS.
8.52 Warning# 4 Pure Water Pressure Above Requirement
Condition
If Pure Water Line option is present: Pure water pressure measured by flowmeter is above requirements for 5
consecutive seconds.
General Troubleshooting Additional Information
Pure water pressure is too high.
1. Contact STERIS.