8 Troubleshooting
194
920515770_J
Vision® 1300 Series
Operator Manual
8.88 Recirculation pump abnormally noisy
Condition
Recirculation pump abnormally noisy.
General Troubleshooting Additional Information
Debris screen clogged.
1. Remove and clean debris screen (refer to Section
7.4, Debris Screen).
2. If situation reoccurs, contact STERIS.
8.89 Touch screen does not work properly
Condition
Touch screen does not work properly.
General Troubleshooting
Additional Information
Washer/disinfector needs to be restarted.
1. Restart washer/disinfector using Power (ON/OFF)
toggle switch.
Touch screen lost calibration or has been damaged.
1. Contact STERIS.
8.90 Travelers moving too slowly
Condition
Travelers moving too slowly.
General Troubleshooting Additional Information
Traveler cable(s) and pulley(s) malfunction or damage.
1. Contact STERIS.
8.91 Travelers vibrating
Condition
Travelers vibrating.
General Troubleshooting Additional Information
Traveler cable(s) and pulley(s) malfunction.
1. Contact STERIS.