EasyManua.ls Logo

Steris Vision 1321 - Bar code reader (accessory) does not operate properly; Chamber does not drain properly; Chemical leaks from mechanical core dosing system

Steris Vision 1321
210 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8 Troubleshooting
173
Operator Manual
Vision® 1300 Series
920515770_J
8.34 Alarm# 32 Chemical #2 Low Level
Condition
Chemical #2 low-level sensor does not send a signal to control.
Alarm is triggered by...
"Send chemical #2 level" value from dosing system = 0.
NOTE: This alarm is only monitored in Ready mode.
General Troubleshooting
Additional Information
Chemical pickup tube is out of container.
1. Press ALARM REPLY on touch screen to stop alarm
buzzer and acknowledge displayed alarm message.
2. Abort cycle.
3. Verify if chemical #2 pickup tube is out of
container.
4. If so, insert pickup tube into chemical container.
5. If condition reoccurs, contact STERIS.
If chemical #2 pickup tube is in chemical container,
chemical level is too low or container is empty.
1. Let cycle proceed through all cycle phases.
2. When cycle is completed, replace empty container
with new one.
3. If condition reoccurs, contact STERIS.
8.35 Alarm# 33 Chemical #3 Low Level
Condition
Chemical #3 low-level sensor does not send a signal to control.
Alarm is triggered by...
"Send chemical #3 level" value from dosing system = 0.
NOTE: This alarm is only monitored in Ready mode.
General Troubleshooting Additional Information
Chemical pickup tube is out of container.
1. Press ALARM REPLY on touch screen to stop alarm
buzzer and acknowledge displayed alarm message.
2. Abort cycle.
3. Verify if chemical #3 pickup tube is out of
container.
4. If so, insert pickup tube into chemical container.
5. If condition reoccurs, contact STERIS.

Table of Contents

Related product manuals