7 — Routine Maintenance
145
Operator Manual
Vision® 1300 Series
920515770_J
14. Enter predefined Service login password using available touch pads on keypad, then press
to confirm
entry.
Display shows Opening door screen.
15. Reinstall four previously removed clips and spray arm plugs on each end of traveler spray arms.
16. Carefully inspect all jets along traveler spray arms and remove any visible debris by pulling them out of the jets.
17. Remove debris screen and rinse off under running water and reinstall (see Section 7.4, Debris Screen).