5 Operating Instructions
74
920515770_J
Vision® 1300 Series
Operator Manual
.
... and display shows:
2. If desired, press Silence buzzer on touch screen to silence buzzer.
IMPORTANT: If Emergency Stop pushbutton is pressed before the Thermal Rinse phase is completed, a supervisor may
need to enter login name and corresponding password to open load door because thermal disinfection was not
performed. If password protection is activated, refer to Section 6.6, Logging in Supervisor Mode.
3. If required, doors can then be manually pushed open sideways to access washer/disinfector chamber.
4. Once situation is resolved, pull Emergency Stop pushbutton.
.
5. Press Alarm reply on touch screen to acknowledge alarm. This records time at which operator acknowledged the
alarm. Display shows: