6 — Cycle and Control Value Programming
81
Operator Manual
Vision® 1300 Series
920515770_J
WARNING
INEFFECTIVE PROCESS HAZARD
Customer is responsible for validating the performances of their own customized cycles. STERIS does not
recommend setting cycle parameters below those validated cycles.
Cycles include various sequences of phases:
• Pre-Wash(INSTRUMENT cycle only - option)
• Thermal Rinse
• Wash
• Drying
Each phase has time and/or temperature values, some of which can be modified in Supervisor mode within a range
when creating customized cycles. To help ensure safe and effective operation, some preset values cannot be modified.
NOTE: See Table 6-1, Cycle Description Chart, for more details on phase specifications for cycles using Prolystica® Ultra
Concentrate chemicals or cycles using other chemicals.