4
Life Science Applications
Nucleic Acid Labels
152 Evolution 200 Series User Guide Thermo Scientific
Related Topics
Measurement Tab for Nucleic Acid Labels
Instrument Tab for Nucleic Acid Labels
Accessories Tab for Nucleic Acid Labels
Samples Tab for Nucleic Acid Labels
Measurement Tab for Nucleic Acid Labels
Click Settings in Nucleic Acid Labels to display the
Measurement tab in the right pane.
Feature Description
Analysis wavelength Wavelength(s) to use for the analysis.
Correction Specifies bichromatic normalization of the absorbance data.
To use uncorrected data, select None. Without baseline correction,
spectra may be offset from the baseline. If this offset is significant, the
calculated protein concentration may be higher than the true value.
Single point subtracts the sample absorbance at the entered
wavelength from the absorbance at every data point.
Sloping baseline draws a line between the data points at the entered
wavelengths, finds its absorbance value at the analytical wavelength,
and subtracts it from the measured absorbance.
Component name Component to quantify.