17
Mercury Lamp
276 Evolution 200 Series User Guide Thermo Scientific
Y To install the mercury lamp
1. Plug the lamp cable into the Hg Lamp connector on the back of the spectrophotometer.
See Connectors.
Languages
2. Follow these instructions to run the mercury lamp Performance Verification (PV) tests or
to perform a wavelength calibration.
To run a PV test, in INSIGHT software click Home and then click Performance
Verification. If you have a custom button configuration, click Home, select the
Classic group and then click Performance Verification.
Select the following PV tests (deselect all other tests) and then click Run:
Wavelength Accuracy - Hg Lines
Wavelength Reproducibility - Hg 546
ADVERTENCIA Evite el riesgo de radiación ultravioleta en el compartimento de la
lámpara de mercurio. Proteja sus ojos y la piel de la exposición.
WARNING Avoid defeating the safety interlock. Keep magnets and strong magnetic
fields away from the mercury lamp when it is installed and the instrument is running.
AVERTISSEMENT Éviter de déjouer le verrouillage de sécurité. Tenez les aimants et
autres champs magnétiques puissants éloignés de la lampe à vapeur de mercure
lorsque celle-ci est installée et que l’instrument est en marche.
WARNUNG Der Sicherheitsschalter darf nicht deaktiviert werden. Magneten und
starke Magnetfelder müssen von der Quecksilberlampe fern gehalten werden, wenn
die Lampe installiert ist und das Instrument läuft.
AVVERTENZA Evitare la disattivazione del dispositivo di sicurezza. Tenere i magneti e
i forti campi magnetici lontano dalla lampada al mercurio mentre è installata e lo
strumento è in funzione.
ADVERTENCIA Evite neutralizar el sistema de interrupción de seguridad. Mantenga
los imanes y campos magnéticos intensos lejos de la lámpara de mercurio cuando esté
instalada y el instrumento en funcionamiento.