TPS-1500W Sealed Peltier Recirculator
Thermo Scientific Evolution 200 Series User Guide 309
Y If an alarm occurs
1. Determine the cause of the alarm by consulting the recirculator front panel display.
2. Press the right (Reset) button, if available, on the recirculator front panel.
See Manual reset prealarm in the Parameters Menu to configure the Reset option to
display in Alarm Mode.
3. Turn off the power switch on the recirculator back panel.
4. Fix the problem. See Water Level Failure and Other System Failures.
5. Turn on the recirculator power switch.
6. If the alarm occurs again, turn off the power switch and contact technical support. See
Contacting Us.
Related Topics
Water Level Failure
Other System Failures
Preinstallation
Installation
Operation
Maintenance
Water Level Failure
The accessory is equipped with a water level sensor that detects small leaks.
Y If a water level failure occurs
1. Press the Reset button, if available, on the recirculator front panel.
See Manual reset prealarm in the Parameters Menu to configure the Reset option to
display in Alarm Mode.
2. Turn off the power switch on the recirculator back panel.
3. Open the sample compartment and clean up any liquids that may have leaked from the
accessory.
4. Tighten all connectors to and from the recirculator.
5. Fill the reservoir. See Filling the Reservoir.
6. Turn on the recirculator power switch.