25
Smart Linear 8-Cell Changer
358 Evolution 200 Series User Guide Thermo Scientific
Installation
Maintenance
Troubleshooting
Calibrating the Cell Changer
The linear cell changer must be calibrated the first time it is installed in a spectrophotometer.
Calibration aligns each cell with the light beam. This is especially important when using
microcells with the cell changer. The calibration settings are stored in the instrument.
Languages
Y To calibrate the linear cell changer
1. In INSIGHT software choose System Settings > Calibrations.
2. In the Linear Cell Changer group, click Calibrate.
A status box shows several intensity values collected on each side of the expected best
location for each cell. The average of the highest readings is stored for each cell.
A message box appears when calibration is completed.
3. Click OK.
CAUTION Avoid pinch hazard. Keep hands and objects clear of the accessory during oper-
ation.
Remarque Évitez les risques de pincement. Éloignez les mains et les objets de l’accessoire
pendant son fonctionnement.
Anmerkung Quetschungen sind zu vermeiden. Hände und Objekte sind während des
Betriebs vom Zubehör fern zu halten
N.B. Evitare il rischio di pizzicamento.Tenere mani e oggetti lontano dall’accessorio
durante l’operazione.
Nota Evite riesgos de atrapamiento.Mantenga manos y objetos alejados del accesorio
durante el funcionamiento.