27
Smart Rotary Cell Changer
398 Evolution 200 Series User Guide Thermo Scientific
6. Close the sample compartment door.
7. Click OK to initialize the accessory.
Initialization reads information about the accessory and moves it to position 1.
8. To use a temperature probe with the cell changer, see “Temperature Probe”.
9. To connect a fluid recirculator or other device to control sample temperature on a
thermostatted cell changer, see “Installing a Thermostatted Accessory.”
10. Calibrate the cell changer (linear cell changers only) and set the software parameters to
display the cell changer status monitor and configure the accessory for operation.
Related Topics
Opening and Closing the Sample and Reference Compartment Doors
Removable Panels
Installing a Smart Accessory
Installing a Liquid Thermostatted Accessory
Temperature Probe
Replacing the Cell Holder
Installation
Replacing the Cell Holder
Initialize the accessory after installing a cell holder.
Y To install a cell holder
1. Line up the tab on the underside of the cell holder with the notch on the motor shaft.
2. Place the cell holder on the shaft.
Tab on
underside of
cell holder
Notch on motor
shaft of Smart base
unit