28
Smart Sipper
Thermo Scientific Evolution 200 Series User Guide 421
Y To clean the accessory exterior
1. Remove the accessory from the sample compartment. See Installing a Smart Accessory.
2. Use a lint-free cloth dampened with plain water to wipe the area near the cell holder.
3. Dry the surface with another cloth.
Related Topics
Routine Maintenance
Replacing Pump Tubing
Cleaning Flow Cells
Replacing Flow Cell Tubing
Maintenance
Replacing Pump Tubing
Examine pump tubing weekly and when pumping efficiency is impaired. Replace crimped or
worn tubing (see Installing Pump Tubing).
If you install a different type or size tubing, recalibrate before analyzing samples (see
Calibration).
Related Topics
Routine Maintenance
Cleaning the Accessory
Cleaning Flow Cells
Replacing Flow Cell Tubing
Maintenance
Cleaning Flow Cells
Rinse the flow cell and inlet tubing thoroughly with laboratory grade purified water after each
analysis and before turning off the Smart Sipper. Keep the cell filled with purified water when
not in use.
At least once a week, remove the flow cell from the accessory and soak it in a solution of
Decon™ detergent (or equivalent), diluted according to the manufacturers instructions. When
finished, reconnect the cell and flush it thoroughly with laboratory grade purified water.