29
Temperature Probe
Thermo Scientific Evolution 200 Series User Guide 439
2. Use tweezers or a pointed tool to remove the small O-ring from the end of the liner.
3. Press the new O-ring fully into the liner.
The O-ring must remain in place when you turn the liner upside down.
4. Use the screw to insert the liner back into the stopper (press to secure the O-ring).
5. Remove the screw from the liner.
Y To replace the gaskets on the square stopper
1. Use tweezers or a pointed tool to remove the gaskets.
2. Stretch each new gasket onto the stopper and press it into the groove.
Related Topics
Routine Maintenance
Cleaning Temperature Probes
Maintenance
Screw
Liner
O-ring
Gaskets