8-12 Imaging Subsystem
Imaging Subsystem Components 5DX Series 3
Figure 8-6: Deflection Signals
The magnitude of the deflection is defined by the values of the FOVs look-up
table loaded into the X-ray Scan Controllers memory. The X-ray Scan Controller
contains four memory banks: Bank 0, Bank 1, Bank 2 and Bank 3. The banks are
look-up tables containing information relating to a Field of View (FOV). Bank 0 is
dedicated to containing the information for the Dump Ring. Banks 1 through 3 are
loaded as required to contain information about specific FOVs.
Y-A xis
X-Axis
Quadrant 1
;$[LV3RVLWLYH
<$[LV3RVLWLYH
Quadrant 2
;$[LV3RVLWLYH
<$[LV1HJDWLYH
Quadrant 4
;$[LV1HJDWLYH
<$[LV3RVLWLYH
Quadrant 3
;$[LV1HJDWLYH
<$[LV1HJDWLYH
Y-A xis
X-Axis
3RVLWLYH
1HJDWLYH
3RVLWLYH
1HJDWLYH