8-18 Imaging Subsystem
Imaging Subsystem Components 5DX Series 3
Focus Coil The Focus Coil electromagnetically concentrates the electron
beam before it strikes the X-ray Target. The Focus Coil receives power from
the Focus Coil Power Supply located in the Electronics Bay of the 5DX
System. The Focus Coil Power Supply is a precision current-controlled power
source that supplies the Focus Coil with 15-20 volts dc at 1-3 amperes of
conditioned power. The Focus Coil Power Supply is set at the factory and
should not be adjusted.
Deflection Coil The deflection coil electromagnetically steers the electron
beam. The deflection coil is actually a pair of directional (X and Y) coils that
cause the electron beam to follow a circular path around the X-ray Target.
X-ray Target The X-ray Target is a thin tungsten substrate that is attached
to a base plate. As the electrons strike the X-ray Target, electrons of the
tungsten atoms are displaced causing the emission of high energy photons (X-
rays). Additional high energy X-rays are generated when the electrons collide
with the nucleus of molecules thereby abruptly decelerating and causing a
Bremsstrahlung Effect.
Dump Ring The dump ring is used during automatic test as a means to
quickly terminate X-ray flux. When the electron beam hits the dump ring, X-
rays are blocked. This greatly reduces the radiation dose to customer panel
reference point components. (See Figure 8-7.)
X-ray Tube Electronics
The X-ray Tube Electronics assembly contains one serial port, daisy chained
with the X-ray High Voltage (HV) Power Supply. The X-ray HV Power Supply
supplies the X-ray Tube Electronics with -160 kilovolts dc, +15 volts dc and -15
volts dc. The X-ray Tube Electronics provides filament drive control used to make
precise X-ray Tube parameter adjustments. These adjustments are accomplished
by regulating the voltage potentials applied to the X-ray Tube filaments (the
Control Grid, and the Focus Grid). The maximum electron beam current for the X-
ray Tube is factory set to 0.100 mA (±0.010mA). To keep the X-ray Tube
Electronics cool, a fan is mounted in the top of the X-ray Tube Tower. The fan
receives power from the -15 volts dc isolated supply.

Table of Contents

Other manuals for Agilent Technologies 5DX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5DX and is the answer not in the manual?

Agilent Technologies 5DX Specifications

General IconGeneral
BrandAgilent Technologies
Model5DX
CategoryMedical Equipment
LanguageEnglish

Related product manuals