Agilent 5DX Service Guide 3-27
5DX Series 3 Operator Control Console
18. Push the Plunger so it is flush with the Secondary Gap Adjuster as shown in
Figure 3-21.
Figure 3-21: Secondary Gap Adjuster Placement Using Lower Slot
19. Verify that the ohmmeter reads zero resistance.
20. Using a T-10 Torx driver, tighten the two screws on the Plunger.
21. Insert the higher slot of the Secondary Gap Adjuster between the Switch and
the Plunger as shown in Figure 3-22. To do this, push up on the Outer Barrier
and insert the Secondary Gap Adjuster between the Switch and the Plunger.