MEMory Subsystem 6
U2000 Series Programming Guide 153
Syntax
Parameters
Example
Query
MEMory:TABLe:GAIN[:MAGNitude]?
The query returns a list of offset points for the currently selected table.
Item Description/Default Range of Values
numeric_value A numeric value for the calibration/ offset
factors. The units are PCT.
1.0 to 150.0
MEM:TABL:SEL "Sensor_1"
MEM:TABL:GAIN 97,99.5,97.4
This command enters a reference offset
factor of 97 %, 99.5 % and 97.4 % into the
sensor frequency dependent offset table.
MEM
:TABL
:GAIN
:MAGN
numeric_value
Space
,
?