7SENSe Subsystem
168 U2000 Series Programming Guide
Query
SENSe[1]:AVERage:COUNt:AUTO?
The query enters a 1 or 0 into the output buffer indicating whether automatic
filter length is enabled or disabled.
1 is returned when automatic filter length is enabled
0 is returned when automatic filter length is disabled
Query Example
Error Messages
If SENSe[1]:AVERage:COUNt:AUTO is set to ON while SENSe[1]:MRATe is set
to FAST, the error –221, “Settings Conflict” occurs. However, automatic
averaging is enabled but the SENSe[1]:AVERage:STATe command is not
automatically set ON.
AVER:COUN:AUTO? This command queries whether automatic
filter length selection is on or off for
power sensor.