Section 8-3 - Software Loading Procedure 8-125
8-3-5-3 Printer setting Back-up
Before starting upgrade, write down, print out or save the printer setting for back-up.
NOTE: All the customized image setting of digital printer will be lost after software loading. Take the
back up and restore all the digital printers setting at software loading.
NOTE: This printer setting is for only digital printers, not analog printers.
1.) Insert the CD or removable media and Format.
2.) Go to Utility -> System -> Peripherals and click the Printers.
3.) Press the left button of the trackball on the printer which need to be backup -> Click “properties”.
Figure 8-168 Printer set backup
Figure 8-169 Printer set backup

Table of Contents

Other manuals for GE LOGIQ P5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P5 and is the answer not in the manual?

GE LOGIQ P5 Specifications

General IconGeneral
Power Supply100-240VAC, 50/60Hz
Display15" LCD
Doppler ModesColor Doppler, Power Doppler
Imaging ModesB-Mode, M-Mode
ProbesConvex, Linear, Phased Array
ApplicationsAbdominal, Vascular, Cardiac, OB/GYN, Small Parts, Musculoskeletal
ConnectivityDICOM, USB, Ethernet
Probe Connectors3
PortabilityMobile
Operating Temperature10°C to 40°C
Relative Humidity30%

Related product manuals