EasyManua.ls Logo

GE LOGIQ P5

GE LOGIQ P5
532 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
8-134 Section 8-4 - Software Installation
8-4-1 Installing Base Image Software (cont’d)
8.) Press Enter key and then the following screen will be shown. (If you want to abort installation
procedures, press CTRL+C.)
9.) Press
A key to install the complete disk.
Figure 8-185 Base Image Software Installation Menu
NOTICE
Select A when installing the new HDD or upgrading from BT09. (HDD capacity should be minimum
120GB)
Select
B when replacing C:\ partition only. It means all customer data are NOT deleted. The Ghost pro-
cedure is performed with data of the drivers D, E, F and Z left.
Select
C when replacing C:\ partition and Z:\ partition. It means all customer data are NOT deleted. The
Ghost procedure is performed with data of the drivers D, E and F left. Z:\ partition used for reloading
C:\ partition and Application software will be replaced.
Drive Partition Usage
(GRUB) grub data
C System Files and Application
D User Defined and Preset
E Patient Archive data
FSwap
Z C:\partition Repository for Reloading Function
Table 8-54 Drive Partition

Table of Contents

Other manuals for GE LOGIQ P5

Related product manuals