DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
Section 8-4 - Software Installation 8-135
8-4-1 Installing Base Image Software (cont’d)
10.)Approx. twenty(20) minutes later, the following screen appears. Eject the Base Image Software
Disk from the CD/DVD ROM drive.
11.)After removing the disk from the drive, Press ALT + CTRL + DEL key.
NOTE: After Installing Base Image Software, the “General failure reading drive A. Abort, Retry, Fail?”
message can be appeared, but Ignore it! All procedure would be successfully done.
12.)The following message box can be displayed after the SYSTEM boot up. Press Yes.
Figure 8-186 Installation Completion
WARNINGWARNING
Remove the disk from the drive. Otherwise you will be encountered the Base Image
Software installation screen iteratively.
Figure 8-187 System Setting Change

Table of Contents

Other manuals for GE LOGIQ P5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P5 and is the answer not in the manual?

GE LOGIQ P5 Specifications

General IconGeneral
Power Supply100-240VAC, 50/60Hz
Display15" LCD
Doppler ModesColor Doppler, Power Doppler
Imaging ModesB-Mode, M-Mode
ProbesConvex, Linear, Phased Array
ApplicationsAbdominal, Vascular, Cardiac, OB/GYN, Small Parts, Musculoskeletal
ConnectivityDICOM, USB, Ethernet
Probe Connectors3
PortabilityMobile
Operating Temperature10°C to 40°C
Relative Humidity30%

Related product manuals