DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
8-176 Section 8-6 - Software Reloading
8-6-2-2 Software Reloading (cont’d)
11.)Press Time Zone
12.)Change the time zone to the local time zone from Table 1.
13.)Press “Apply”
14.)Press “Date & Time” tab.
Figure 8-243 Time Zone
Figure 8-244 Date & Time