Configure the MUSE System for Network Communication
13. In the End Point Configuration area of the window, select the protocol(s) you are
using to communicate with MUSEAPI3 and enter the port value(s).
You must have at least one protocol enabled, and you may have more than one.
If any protocols are selected that you do not want, remove the check mark.
You are advised to use the values below for the ports:
Protocol Recommended Port Values
HTTP 8100
HTTPS The port for SSL, as configured by the customer.
net.tcp 8101
NOTE:
For more information on the available communication protocols, see Table
116: MUSEAPI3 Installation Prerequisites on page 312.
14. Determine whether you want to check the box next to PID Search: Combined
Test Lists to change the Patient Conflict behavior of the MUSEAPI3 and do one of
the steps below:
• Select to enable the PID Search: Combined Test List. When performing
a Patient ID search while this option is enabled, MUSEAPI3 automatically
combines all tests for that Patient ID for the same MUSE site, even if there is a
Patient ID/Last Name mismatch.
• To disable the PID Search: Combined Test List, leave the box unselected.
When performing a Patient ID search while this option is disabled, MUSEAPI3
includes patient conflicts if there is a Patient ID/Last Name mismatch within
the same site.
This setting can always be changed later.
NOTE:
MUSEAPI v3.1 handles patient conflicts within the same MUSE site differently
than MUSEAPI v3.0. MUSEAPI v3.1 only provides a response that includes