OPERATING DESCRIPTION
87
Von Latch: ON
time
time
load
Output
PEL-2000A
series load
V
I
Von Latch: OFF
time
time
load
Output
PEL-2000A
series load
V
I
Von
Von
As can be seen in the diagram below, when Von-Latch
is set to off, the load module will start to sink current
when the Von-voltage limit has been tripped. It will
stop sinking current when the output drops below the
Von voltage limit.
Von Latch: OFF - Start
time
Von Latch: OFF - Stop
Output
time
time
PEL-2000A
series load
time
V
I
load
load
Von
Output
PEL-2000A
series load
V
I
CH CONT
Channel Control. When Channel control is
activated (External) it can be used to monitor the
voltage and current output of the load as well as
turn loads on or off remotely from the Channel
Control (CH CONT) connectors located on the rear
panel.
For more information about channel control, see
external voltage control on page 89.
Independent
The Independent setting will allow the load
modules to be controlled independently from the
mainframe.