PEL-2000A Series User Manual
88
Load Delay time is used to delay activating a load
(up to 10 seconds) after the load key has been
pressed. However the Load D-Time setting will
only work for loads that are initiated manually or
when the PEL-2000A series mainframe is
configured to Auto load (page 210) at run time.
The Response setting sets the bandwidth of the
load to 1kHz (normal) or 100kHz (fast). The
Response setting is particularly important for
limiting startup current.
The current, resistance, voltage and power setting
can have the step resolution configured for each
channel. The step resolution refers to the step
resolution of the coarse adjustment of these settings.
The fine adjustment cannot be configured, see page
197 for details.
For example if the step resolution for CCH (CC
high range) is .5 A, then the resolution can be
incremented in .5A steps;
8.08.59.09.5
The step resolution parameters apply to the
following:
CCH Step – CC high range
CCL Step – CC low range
CRH Step – CR high range
CRL Step – CR low range
CVH Step – CV high range
CVL Step – CV low range
CPH Step - CP high range
CPL Step - CP low range