System
Interface
MGCplus A0534-30.0 HBM: public 277
Direct Ethernet connection between PC and MGCplus with permanent
IP addresses
For example, if your computer has the IP address 172.34.24.13, your
MGCplus should have IP address 172.34.24.x (x13, x255) if you are using
a subnet mask 255.255.255.0.
If you connect the MGCplus to a PC via a hub, use a patch cable.
The Ethernet interfaces of the CP52 are both DHCP- and APIPA-capable
and have this setting when delivered from the factory.
The Ethernet interfaces of the CP22 and CP42 are not DHCP- or
APIPA-capable and are therefore delivered with the following permanent IP
addresses:
CP42:
IP address = 192.168.169.134
Subnet mask = 255.255.255.0
Router = 0.0.0.0
CP22:
IP address = 192.168.169.252
Subnet mask = 255.255.255.0
Router = 0.0.0.0