EasyManua.ls Logo

Hioki IM3536 - 5.7 Correcting Measured Values with a User-specified Correction Coefficient (Correlation Correction)

Hioki IM3536
276 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
115
Load Correction (Correcting Values to Match Reference Values)
Setting the correction frequency
1
Touch the FREQ key.
2
Enter the correction frequency with the
numeric keypad, and touch the unit key
to conrm the setting.
2
Touch when you
wish to perform load
correction during DC
measurement.
1
Touch when you wish
to cancel input. (This
dialog box will be
closed.)
Settable range: DC, 4 Hz to 8 MHz
*
*: The maximum frequency varies with the
cable length (p. 215).
If you make a mistake, touch the C key to
reenter the value.
3
Touch the EXIT key.
The dialog box will be closed.
Selecting the correction range
1
Touch the RANGE key.
2
Select the range for correction.
Touch when you wish to enable LOW Z op-
eration.
3
Touch the EXIT key.
The dialog box will be closed.
The selection of available ranges will vary with the correction frequency. For more information, see p. 211