Appendix B. Processing Flags
Code Description Outcome
MFM Magnet of the Magnetic
Wash Station cannot be
positioned.
MTU is invalid.
Contact Hologic Technical Support.
MHF MTU aspiration failure. Sample is invalid.
Contact Hologic Technical Support.
ML1 The Magnetic Wash
Station reported the
presence of fluid at the
residual position after
aspiration.
Sample is invalid.
Contact Hologic Technical Support.
ML2 The Magnetic Wash
Station reported a liquid
level sense too high
after Wash dispense.
Sample is invalid.
Run the Mag Wash Clean Procedure.
Samples must be retested. Contact Hologic
Technical Support.
ML3 The Magnetic Wash
Station reported a liquid
level sense too low after
Wash dispense.
Sample is invalid.
Check Wash connections. Contact Hologic
Technical Support.
ML5 The Magnetic Wash
Station reported a liquid
level sense too high
after final aspiration.
Sample is invalid.
Run the Mag Wash Clean Procedure.
Samples must be retested. Contact Hologic
Technical Support.
MMF Mag Wash Mixer
movement failure
MTU is invalid.
Contact Hologic Technical Support.
MT1 One or more tiplets were
not detected on the
Magnetic Wash Station
aspirators at the start of
the wash cycle. This
may be due to a
hardware problem or
defective MTU.
MTU is invalid.
Samples must be retested. Contact Hologic
Technical Support if the problem is chronic.
Ensure there are tiplets present in MTUs
before loading.
Table 116. Assay Processing Flags (Continued)