Appendix B. Processing Flags
Code Description Outcome
MT2 One or more tiplets were
not detected on the
Magnetic Wash Station
aspirators after the wash
cycle was complete.
This may be due to a
hardware problem.
MTU is invalid.
Contact Hologic Technical Support.
MOS Mag Wash station taken
out of service.
Run the Mag Wash Clean Procedure.
Contact Hologic Technical Support.
NTI The sample was not
tested due to a
hardware error.
Sample is invalid.
Contact Hologic Technical Support.
NTP No pipettor tips
available.
Reload pipettor tips.
Contact Hologic Technical Support if pipettor
tips are still available in the Tip Drawers
when this flag was generated.
PCL1 Process control failure
for Auto Detect 1.
Sample is invalid.
Contact Hologic Technical Support.
PCL2 Process control failure
for Auto Detect 2.
Sample is invalid.
Contact Hologic Technical Support.
PEO The replicate was tested
using an assay installed
as Performance
Evaluation Only.
Informational message only. No action
necessary.
PFR Pump failure during
reagent pipetting.
Sample is invalid.
Contact Hologic Technical Support.
PFS A pump failure occurred
during sample pipetting.
Sample is not tested. Sample results marked
as Error.
Contact Hologic Technical Support.
PFT A pump failure occurred
during TCR pipetting.
If possible, system will retry to pipette TCR
into MTU. If there is a hardware failure,
contact Hologic Technical Support.