Panther System
®
Operator’s Manual
AW-17791-001 Rev. 002 (EN)
278
Appendix B. Processing Flags
Code Description Outcome
PMFR Pipettor movement
failure during reagent
pipetting.
Sample is invalid.
Contact Hologic Technical Support.
PMFS Pipettor movement
failure during specimen
pipetting.
If possible, system will retry to pipette TCR
into MTU. If there is a hardware failure,
contact Hologic Technical Support.
PMFT Pipettor movement
failure during TCR
pipetting.
If possible, system will retry to pipette TCR
into MTU. If there is a hardware failure,
contact Hologic Technical Support.
PTF Failure to pick a tip. Sample is invalid.
Sample must be retested. Contact Hologic
Technical Support if problem is chronic.
QNS Sample quantity not
sufficient for processing.
Sample is not tested.
Sample must be retested, or re-collected if
volume is too low.
Check all samples for adequate volume prior
to loading.
RBEA Amplification reagent
bottle empty.
Assay reagent kit is invalid due to low volume
in specified reagent bottle. Any tests in
process with this kit are invalid. Retest
invalidated samples with a new kit. Load a
new kit to test additional samples.
RBEE Enzyme reagent bottle
empty.
Assay reagent kit is invalid due to low volume
in specified reagent bottle. Any tests in
process with this kit are invalid. Retest
invalidated samples with a new kit. Load a
new kit to test additional samples.
RBEP Probe reagent bottle
empty.
Assay reagent kit is invalid due to low volume
in specified reagent bottle. Any tests in
process with this kit are invalid. Retest
invalidated samples with a new kit. Load a
new kit to test additional samples.