Appendix B. Processing Flags
Code Description Outcome
RBEPM Promoter reagent bottle
empty.
Assay reagent kit is invalid due to low volume
in specified reagent bottle. Any tests in
process with this kit are invalid. Retest
invalidated samples with a new kit. Load a
new kit to test additional samples.
RBESL Selection reagent bottle
empty.
Assay reagent kit is invalid due to low volume
in specified reagent bottle. Any tests in
process with this kit are invalid. Retest
invalidated samples with a new kit. Load a
new kit to test additional samples
RBET TCR reagent bottle
empty.
Assay reagent kit is invalid due to low volume
in specified reagent bottle. Any tests in
process with this kit are invalid. Retest
invalidated samples with a new kit. Load a
new kit to test additional samples.
RBETR TER reagent bottle
empty.
Assay reagent kit is invalid due to low volume
in specified reagent bottle. Any tests in
process with this kit are invalid. Retest
invalidated samples with a new kit. Load a
new kit to test additional samples.
RDFA RDV indicated a failure
during dispense of
Amplification reagent.
MTU is invalid.
Contact Hologic Technical Support.
RDFE RDV indicated a failure
during dispense of
Enzyme reagent.
MTU is invalid.
Contact Hologic Technical Support.
RDFP RDV indicated a failure
during dispense of
Probe reagent.
MTU is invalid.
Contact Hologic Technical Support.
RDFPM RDV indicated a failure
during dispensing of
Promoter reagent.
MTU is invalid.
Contact Hologic Technical Support.