Panther System
®
Operator’s Manual
AW-17791-001 Rev. 002 (EN)
280
Appendix B. Processing Flags
Code Description Outcome
RDFS Sample dispense
verification failed. Once
flagged with an RDFS
error, the sample is not
processed further.
Sample is not tested. Sample results marked
as Invalid.
Ensure that there are no interfering
substances and adequate volume in the
sample tube. Sample must be retested, or
recollected if volume is low.
RDFSE Sample eluate dispense
verification fails.
Sample is not tested. Sample results marked
as Invalid. Retest invalidated sample.
RDFSL RDV indicated a failure
during dispense of
Selection reagent.
MTU is invalid.
Contact Hologic Technical Support.
RDFT Reagent dispense
failure during dispense
of TCR reagent.
If possible, system will retry to pipette TCR
into MTU. If there is a hardware failure,
contact Hologic Technical Support.
RDFTR Reagent dispense
failure during dispense
of TER reagent.
MTU is invalid.
Contact Hologic Technical Support.
RT Duplicate Sample ID. Informational Message only. No action
necessary.
RTL Reagent pipettor tip
loss.
MTU is invalid.
Contact Hologic Technical Support.
RUO The replicate was tested
using an assay installed
as Research Use Only.
Informational Message only. No action
necessary.
RVHA Amplification reagent
bottle above the
expected volume.
Ensure that reagent bottles are seated
properly.
Check Amplification reagent bottle for correct
volume. Contact Hologic Technical Support.
RVHE Enzyme reagent bottle
above the expected
volume.
Ensure that reagent bottles are seated
properly.
Check Enzyme reagent bottle for correct
volume. Contact Hologic Technical Support.
Table 116. Assay Processing Flags (Continued)