Panther System
®
Operator’s Manual
AW-17791-001 Rev. 002 (EN)
281
Appendix B. Processing Flags
Code Description Outcome
RVHP Probe reagent bottle
above the expected
volume.
Ensure that reagent bottles are seated
properly.
Check Probe reagent bottle for correct
volume. Contact Hologic Technical Support.
RVHPM Promoter reagent bottle
above the expected
volume.
Ensure that reagent bottles are seated
properly.
Check Promoter reagent bottle for correct
volume.
Contact Hologic Technical Support.
RVHSL Selection reagent bottle
above the expected
volume.
Ensure that reagent bottles are seated
properly.
Check Selection reagent bottle for correct
volume. Contact Hologic Technical Support.
RVHT TCR reagent bottle
above the expected
volume.
Ensure that reagent bottles are seated
properly.
Check TCR reagent bottle for correct volume.
Contact Hologic Technical Support.
RVHTR TER reagent bottle
above the expected
volume.
Ensure that reagent bottles are seated
properly.
Check TER reagent bottle for correct volume.
Contact Hologic Technical Support.
S The Sample barcode ID
was entered by the
operator, using the
handheld scanner.
Informational Message only. No action
necessary.
SID An auto-generated
specimen ID was
automatically assigned
by the System due to an
unreadable barcode.
Informational Message only. Correct the SID
by following the instructions in the Editing
Auto-Generated Sample IDs section.
SRR Sample pipetting
operation aborted.
Sample is invalid.
Contact Hologic Technical Support.
Table 116. Assay Processing Flags (Continued)