Panther System
®
Operator’s Manual
AW-17791-001 Rev. 002 (EN)
381
Appendix C. System Messages
Code Message Possible Cause Solution
406.002.00001 Database restored
from backup {0}
N/A No action necessary.
407.001.00001 User logged on N/A No action necessary.
407.001.00002 User logged off N/A No action necessary.
407.001.00003 User shutdown N/A No action necessary.
407.001.00004 User added: {0} N/A No action necessary.
407.001.00005 User edited: {0} N/A No action necessary.
407.001.00006 User deleted: {0} N/A No action necessary.
407.001.00007 User password
changed
N/A No action necessary.
407.002.00001 Valid Assay
Reagent Kit
Loaded: {0}
N/A No action necessary.
407.002.00002 Assay Reagent Kit
Unloaded: {0}
N/A No action necessary.
407.002.00003 AR Inventory
started
N/A No action necessary.
407.002.00005 Assay Reagent Kit
has expired in lane
{0}.
Identified reagent kit
has expired and
can no longer be
used.
Load a new assay
reagent kit to process
additional tests.
407.002.00006 Load/unload
carousel reagents
and close carousel
door
The system is ready
to load/unload
carousel reagents.
1. Load/unload
reagents in the
carousel according
to current task.
2. Close the TCR
Door completely
before continuing.
Table 118. System Messages (Continued)