Panther System
®
Operator’s Manual
AW-17791-001 Rev. 002 (EN)
382
Appendix C. System Messages
Code Message Possible Cause Solution
407.002.00007 {0} cutoff set to {1}
for {2} reagent kit in
lane {3}.
Supplemental
assay reagent
information has
been accepted. The
cutoff values for an
assay reagent kit
has been changed.
No action necessary.
407.002.00008 {0} cutoff set to null
for {1} reagent kit in
lane {2}.
N/A No action necessary.
407.003.00001 Universal Fluid
drawer opened
during Assay
Processing.
Universal Fluid
drawer opened
during Assay
Processing.
1. Close the fluid
drawer.
2. Do not open
Universal Fluid
drawer during
Assay Processing.
407.003.00002 Fluids Inventory
started
N/A No action necessary.
407.003.00004 Data Integrity:
Universal Fluid
bottles were
incorrectly marked
as loaded.
Database Integrity
error.
1. Close the fluid
drawer.
2. Contact Hologic
Technical Support.
407.003.00005 Universal Fluid
drawer opened
during Priming.
The Universal
Fluids drawer is
opened during
priming.
No action necessary.
407.004.00001 Test Order added to
Sample {0}: Assay
{1}
N/A No action necessary.
407.004.00002 Test Order on
Sample {0}
updated: Sample
{1}, Analytes {2},
Priority {3}
N/A No action necessary.
Table 118. System Messages (Continued)