Appendix C. System Messages
Code Message Possible Cause Solution
407.004.00003 Test Order deleted
from Sample {2}:
Assay {0}, Analytes
{1}
N/A No action necessary.
407.004.00004 Test Order Analytes
cannot be changed
once Sample has
been pipetted:
Sample {0}, Assay
{1}, Analytes {2}.
Tried to change
Test order Analyte
type after sample
was pipetted.
1. Cannot remove test
orders once the test
order has already
been pipetted.
2. Edit the test orders
for a specimen prior
to pipetting.
407.004.00005 Test Order for
Sample {0} Verified
N/A No action necessary.
407.004.00006 Unable to resolve
assay for test order:
{0} - {1}
System software or
database error.
1. Shutdown and
restart.
2. If the problem
persists, contact
Hologic Technical
Support.
407.004.00007 Attempted to Verify
a Test that was not
complete: {0} - {1} :
{2}
While verifying
each test order, a
check is made to
verify the test order
is completed.
No action necessary.
407.005.00001 Waste Inventory
started
N/A No action necessary.
407.005.00002 Waste drawer error.
Please open and
close the waste
drawer.
The Waste Drawer
has failed to
properly lock.
1. Firmly close the
Waste Drawer.
2. If the problem
persists, contact
Hologic Technical
Support.
Table 118. System Messages (Continued)