Panther System
®
Operator’s Manual
AW-17791-001 Rev. 002 (EN)
388
Appendix C. System Messages
Code Message Possible Cause Solution
407.006.00007 The sample rack in
lane {0} was
rejected because a
1-character
barcode is present
in the rack.
The issue could be
due to one of the
following:
l The sample
barcode had the
incorrect
number of
characters
(acceptable
range: 4–20).
l The scanner
misread the
Tube Position
Barcode or the
sample barcode.
1. Remove the rack,
ensure that the
position and
sample barcodes
are intact and load
the rack again.
2. 4–20 barcode
characters are
required for the
sample barcodes.
3. If the problem
persists, contact
Hologic Technical
Support.
407.006.00008 Sample ID hand
scanned: {0}
The barcode
scanner was used
to enter a new
Sample ID.
No action necessary.
407.006.00009 Sample ID
manually entered:
{0}
The Sample ID was
changed on the
Sample Details
screen.
No action necessary.
407.006.00010 Specimen override
set for sample: {0}
Test order(s) were
added on the
Sample Details
screen.
No action necessary.
407.006.00011 Specimen override
removed for
sample: {0}
Test order(s) were
removed on the
Sample Details
screen.
No action necessary.
407.006.00012 External quality
control type
changed to {0} for
sample: {1}
N/A No action necessary.
407.006.00013 External quality
control removed for
sample: {0}
N/A No action necessary.
Table 118. System Messages (Continued)

Table of Contents