8 Reagents
917 Coulometer
ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿
57
ï¿ at the end of the determination.
Number of determ.
The number of determinations to be carried out with a certain amount of
reagent depends on the type of sample and its amount.
Input range 1 - 999
Selection off
Default value off
Working life
Working life of the reagent.
Input range 1 - 999 days
Selection off
Default value off
Reagent capacity
This parameter is only visible with coulometric reagents.
Water capacity of the reagent.
Input range 1 - 9999 mg
Selection off
Default value off
Drift
This parameter is only visible with coulometric reagents.
During conditioning the measured drift has to be in the following range
for 2 min: 'specified drift value + 50 µg/min'.
Input range 1 - 999 µg/min
Selection off
Default value off
[Status]
Display the status overview of the current values of the reagent monitor-
ing.
[Reagent replacement]
Edit the parameters for the reagent exchange.
"Reagent monitoring / Status" dialog
The current reagent monitoring values are displayed in this dialog.
[Reset]
Reset the values to zero.