8.2 Reagent monitoring
■■■■■■■■■■■■■■■■■■■■■■
58
■■■■■■■■
917 Coulometer
Dialog "Reagent monitoring / Reagent replacement"
The parameters for the reagent replacement are defined in this dialog.
Reagent replacement
The reagent can either be exchanged manually or automatically.
Selection manual | auto
Default value manual
manual
If a monitored parameter has reached the limit set, a message is being
displayed. Then the reagent has to be exchanged manually.
auto
If a monitored parameter has reached the limit set, the method defined
below is started automatically.
Memory
This parameter can only be edited with Reagent replacement = auto.
Memory location the method is loaded from. All memory locations are
selectable, even if they are currently not accessible.
Selection Internal memory | External memory 1 | Exter-
nal memory 2 | Shared memory
Default value Internal memory
Shared memory
Shared directory in the network.
Method
This parameter can only be edited with Reagent replacement = auto.
Method used for emptying the titration cell.
NOTICE
Make sure that the memory is accessible.
Entry 32 characters maximum
Selection Selection of stored methods