Product Specifications Wireless Specification
TM80 Telemetry Monitor Operator’s Manual A - 7
A.7.1.2 Implemented Function
This function is provided to transfer configuration information between the TM80 and to
perform communication between a TM80 and a BP10.
BP10 transmits NIBP parameters, status, and prompt messages to the TM80. The TM80
transmits control information and settings information to the BP10.
A.7.1.3 Function Specification
Data integrity
Code error rate ≤ 1%
Data latency
Total delay of data transmission from BP10 to the TM80 or from
the TM80 to BP10: ≤ 3 second
Total delay of data transmission from the TM80 to the CMS: ≤ 3
seconds
Priority
All communication data type shall have the same priority.
Transmission distance
Distinct vision distance between the TM80s is no less than 3 m
Distinct vision distance between the TM80s and the BP10 is no
less than 3 m.
System capacity
One TM80 can synchronously transport the device configura-
tion to at least other five the TM80s in a 10 m
2
space. Five pairs
of the TM80s and BP10s can communicate in a 10 m
2
space,
with the TM80 communicating with the CMS.
Resistance to wireless interference
The bluetooth function is normal when the distance between
the following interfering devices and the TM80 or BP10 are as
follows:
> 1 m for microwave ovens.
> 0.2 m for cellular communication devices, Wi-Fi
devices, intercoms, and cordless phones.
Communication interruption message
When the bluetooth communication is interrupted, the TM80
displays a prompt message.